8270A and 8370A
Sku: FCAL-8x70A
The 8270A and 8370A are modular systems that can be configured with up to five measurement modules at the same time!
Product Information
A pressure range 100 times smaller than the full scale will meet the control specifications without reducing supply pressure. For example, an 8370A with a 70 MPa (10 000 psi) high range module and a 700 kPa (100 psi) low range module will be able to control pressures below 700 kPa (100 psi) to within ±0.014 kPa (0.002 psi) without reducing the supply pressure.
Flexible configuration options
The 8270A and 8370A are modular systems that can be configured with up to five measurement modules at the same time, so they can handle high pressure, low pressure, and everything in between. Three families of modules provide three levels of performance that let you set up a system matching your needs for accuracy and value.
| Good: PM200 Pressure Measurement Modules |
Better: PM500 Pressure Measurement Modules |
Best: PM600 Measurement Modules |
| 0.02% FS specification makes it ideal for calibrating or testing pressure dial gauges, lower accuracy transmitters, or pressure switches
Rugged silicon pressure sensor design allows for faster control Economic price point helps facilitate the purchase of backup modules, making sure you are never down for calibration |
Highly characterized and linearized silicon pressure sensor provides an economical way of making accurate pressure measurements
0.01% reading measurement uncertainty from 50% to 100% for most ranges allows for wide workload coverage More than 45 ranges, from low pressures up to 20 MPa (3000 psi) to choose from. Your application is covered with this wide flexibility in configuration. |
Fluke Calibration Quartz Reference Pressure Transducer (Q-RPT) technology provides precise measurement with long term stability
0.01% reading measurement uncertainty from 30% to 100% of the modules’ span allows for extremely wide workload coverage Onboard barometer included with absolute mode modules, allowing them to be used in both absolute and gauge mode |
Modern replacements for hydraulic deadweight testers and comparators
Safety: the number one priority
Safety is the top priority in the design and manufacture of our pressure products, especially high-pressure pneumatic controllers. The 8270A and 8370A meet all appropriate safety standards. Each unit is equipped with over-pressure protection as well as a front-panel abort button. Pressing the abort button vents the test pressure and returns the unit to a safe condition. Additionally, an abort valve accessory allows for faster, directed vent, as well as automated venting at loss of power. High-pressure gas does have more stored energy than high-pressure oil, but it is a safe option under the right conditions. By using the 8270A or 8370A with the available accessories, you have safe operation with a clean pressure media.
Calibrate gas devices
The 8270A and 8370A, coupled with the optional CPS, can calibrate gas-filled devices. With its wide pressure range coverage, one controller can accomplish the task that used to require two separate standards.
Easiest to use high-pressure standards available
Automated pressure controllers are some of the simplest, easiest-to-use pressure standards. Controlling and measuring pressure is as simple as typing in the required pressure and pressing Enter. Unlike deadweight testers, you do not need to spin weights at a particular speed or calculate gravity or temperature corrections. Unlike pressure comparators, the process can be fully automated and stability is achieved and maintained in a fraction of the time.
Flexibility to handle a wide variety of applications and workloads
The 8270A and 8370A can be configured with different classes of modules with different pressure ranges. This gives you flexibility to handle a variety of applications and workloads. Uncertainties as good as 0.01 % reading are available across the vast majority of the range, making it an ideal replacement for deadweight testers.
An affordable solution that can grow as your requirements do
With their wide pressure range coverage, the 8270A and 8370A can replace multiple standards at once. Thanks to the modular design, pressure ranges and capabilities can be added at any time, allowing for lower cost of entry in favor of future expansion, growing with your needs.
Ideal for characterisation or calibration in production processes
Whether you manufacture large batches of low cost pressure sensors or customized, expensive pressure sensors for aerospace and other mission critical applications, the 8270A and 8370A are ideal for use in your production line.
Robust design provides unsurpassed control performance
The 8270A and 8370A make use of our unique pressure control technology, first developed to handle some of the harshest pressure calibration applications. This technology provides unsurpassed control performance over a wide pressure range over an extended life cycle, reducing your downtime and maintenance costs.
Modularity makes maintenance a snap
The 8270A and 8370A make use of Fluke Calibration’s modular platform design. Measurement and control modules can be easily removed through the front panel to simplify maintenance or re-ranging requirements.
Control speed meets the demands of a manufacturing environment
With times to setpoint as low as 30 seconds, the 8270A and 8370A can control pressures faster than traditional high-pressure controllers can.
Support for remote communication and automation
The 8270A and 8370A support remote communication over RS-232, GPIB, USB, and ethernet connections. Using either the native SCPI protocol or a variety of emulation modes, these controllers can be used in a wide range of automated processes.
Clean operation to 100 MPa (15 000 psi)
The wide pressure range control allows you to use the same test station for low, medium, or high-pressure work. By using a pneumatic pressure controller in your production process, you can provide your customers with a clean product, free of oil contamination.
Automate with COMPASS® software
COMPASS for Pressure software enables you to automate the 8370A and 8270A and run complete pressure calibration sequences on single or multiple devices under test (DUTs). The 8270A and 8270A feature a full remote interface that also lets you use it with custom software or other data acquisition equipment.
Support and services when you need them
Fluke Calibration offers testing, repair and calibration services to meet your needs quickly and at a fair cost while maintaining the high level of quality that you expect. Our pressure calibration laboratories are accredited for conformance to ISO Guide 17025 and we maintain global calibration and repair facilities.
Get peace of mind and uptime with a Gold CarePlan
A Priority Gold Instrument CarePlan gives you peace of mind and maximum uptime from the people who know your Fluke product best. It includes an expedited annual calibration and extended warranty to help you reduce downtime by a week1 and get the best performance from your instruments. Choose from one-year, three-year or five-year CarePlans. (Note: Priority shipping times vary by country. Contact your local Fluke Calibration sales representative for details.)
Training options help you to get the most from your investment
Fluke Calibration sponsors pressure and flow calibration courses for both novice and expert pressure calibration professionals in our Phoenix, Arizona facility. We also host a series of free webinars about a wide variety of pressure calibration topics. If you need training to help you maintain your fleet of pressure controllers, we can help you there too.
- Automated pneumatic controllers that calibrate a wide range of pressure sensors.
- 8270A: Measures and controls pressures from vacuum to 44 MPa (6000 psi). Configurable with ranges as low as 100 kPa (15 psi).
- 8370A: Measures and controls pressures from atmosphere to 107 MPa (15,000 psi). Configurable with ranges as low as 700 kPa (100 psi).
- Modular design allows for improved maintainability, expandability, and range coverage with control performance of 0.002% of active range
| General specifications | |
| Mains | |
| Power requirements | 100 V ac to 240 V ac, 47 Hz to 63 Hz |
| Fuse | T 2 A, 250 V |
| Max power consumption | 100 W |
| Environment | |
| Operating ambient temperature range | 15 °C to 35 °C |
| Storage temperature | -20 °C to 70 °C |
| Relative humidity | Operating: <80 % to 30 °C, <70 % to 40 °C |
| Storage: <95 %, non-condensing. A power stabilization period of four days may be required after extended storage at high temperature and humidity. | |
| Vibration | MIL-T-28800E |
| Altitude (operation) | <3000 m |
| Warm up time | 15 minutes after power up or module installation, when items previously stored within Operating Ambient Temperature Range. |
| Compliance | |
| Ingress protection | IEC 60529:IP20 |
| Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |
| Weight and Dimensions | |
| Weight (Chassis only) | 13 kg (28.5 lbs) / 15 kg (35.25 lbs) |
| Dimensions | Height: 147 mm (5.78 in) |
| Width: 452 mm (17.79 in) | |
| Depth 488 mm (19.2 in) | |
| Rack mount dimensions | 3U-19 inch rack |
| Electromagnetic Compatibility (EMC) | |
| IEC 61326-1 (Controlled EM environment | IEC 61326-2-1; CIPR 11: Group 1, Class A Group 1 equipment has intentionally generated and/or uses conductivity coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself.Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.Emissions which exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object. The equipment may not meet the immunity requirements of 61326-1 when test leads and/or test probes are connected. |
| USA (FCC) | 47 CFR 15 subpart B, this product is considered an exempt device per clause 15.103. |
| Korea (KCC) | Class A Equipment (Industrial Broadcasting & Communication Equipment) This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes. |
| Communication Interfaces | |
| Primary Remote Interfaces | IEEE, Ethernet, RS-232, USB |
| System Connection | Supports interconnection of 2 or 3 systems |
| Switch Test Connection | Standard BNC Jack: Nominal 24 V dc isolated drive Maximum 30 V dc w.r.t.chassis ground |
| Aux Drivers | 4 external Solenoid Drivers 24 V dc Drive (Maximum drive 6 W continuous per channel |
| Operating characteristics | ||||||||||||||||||||||||||||||
| Control specifications | ||||||||||||||||||||||||||||||
| 95 % of setpoints are within specification limits for stated conditions as calculated by mean plus 1.67 standard deviations of test data. | ||||||||||||||||||||||||||||||
| Control Precision (Dynamic Mode) | 0.002% Range Span or 0.01 kPa (whichever is greater) | |||||||||||||||||||||||||||||
| Control Turndown | 100:1 (typical)
To meet the control specifications, supply pressure should not be greater than 100 times the range of the measurement module. Control turndown is defined as the relationship between the provided supply pressure and the appropriate supply pressure for the range |
|||||||||||||||||||||||||||||
| Low Control Point | 1 kPa (0.15 psi) absolute (8270A only) 7 kPa (1.0 psi) gauge (8370A, 8270A without vacuum pump) |
|||||||||||||||||||||||||||||
| Settling Time | ||||||||||||||||||||||||||||||
| Settling time is the time required to be within 0.005% of setpoint for 10 % steps into volumes up to 50 cm3 for all pressures >7 kPa absolute (8270A) or 7 kPa gauge (8370A). Settling time may be affected by multiple variables, including temperature effects, component flow rates, leaks, and overall volume configurations. | ||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||
| Overshoot (Dynamic A) | 0.08% Full Scale or 2 kPa (whichever is greater) | |||||||||||||||||||||||||||||
| Overshoot (Dynamic B) | 0.008% Full Scale or 2 kPa (whichever is greater) | |||||||||||||||||||||||||||||
| Pressure Limits | ||||||||||||||||||||||||||||||
| Supply Port (8270A/8370A) | Maximum 48 MPa (7 000 psi) gauge / 110 MPa (16 000 psi) gauge Minimum 2 MPa (300 psi) gauge for 8270A and 8370A |
|||||||||||||||||||||||||||||
| Test Port (8270A/8370A) | 44 MPa (6 400 psi) gauge / 107 MPa (15 500 psi) gauge | |||||||||||||||||||||||||||||
| Reference Port | 150 kPa (22 psi) absolute | |||||||||||||||||||||||||||||
| Vent Port | 150 kPa (22 psi) absolute | |||||||||||||||||||||||||||||
| Relief Valves/Rupture Disk | ||||||||||||||||||||||||||||||
| 8270A Chassis Supply port relief valve is set to 52 MPa (7 500 psi)
8370A Chassis Supply has a rupture disk installed which is designed to burst at Modules with full scales ≤ 44 MPa include pressure relief valves |
||||||||||||||||||||||||||||||
| Supply Gas Type | ||||||||||||||||||||||||||||||
| Clean Dry Nitrogen, Helium, Argon, or Air – Industrial Grade Nitrogen, 99.5%+ | ||||||||||||||||||||||||||||||
| Vacuum Supply | ||||||||||||||||||||||||||||||
| >50 liters per minute capacity with Auto Vent feature
System will exhaust gas through the vacuum system when controlling down in pressure. Appropriate protections are necessary. |
||||||||||||||||||||||||||||||
8370A-2 | High-Pressure Modular Controller Chassis, 2 Bay, 100 MPa (15 000 psi)
8370A-5 | High-Pressure Modular Controller Chassis, 5 Bay, 100 MPa (15 000 psi)
8370A-2-PCM | High-Pressure Modular Controller, 2-Bay, W/PCM
8370A-5-PCM | High-Pressure Modular Controller, 5-Bay, W/PCM
8270A-2-BSP | Modular Pressure Controller Chassis, 2 Bay, BSP Manifold
8270A-2-BSP-PCM | Modular Pressure Controller, 2-Bay, BSP, W/PCM
8270A-5-BSP-PCM | Modular Pressure Controller, 5-Bay, BSP, W/PCM
8270A-2-NPT | Modular Pressure Controller Chassis, 2 Bay, NPT Manifold
8270A-2-NPT-PCM | Modular Pressure Controller, 2-Bay, NPT, W/PCM
8270A-5-NPT-PCM | Modular Pressure Controller, 5-Bay, NPT, W/PCM
8270A-2-SAE | Modular Pressure Controller Chassis, 2 Bay, 7/16-20 Manifold
8270A-2-SAE-PCM | Modular Pressure Controller, 2-Bay, SAE, W/PCM
8270A-5-SAE-PCM | Modular Pressure Controller, 5-Bay, SAE, W/PCM
8270A-5-BSP | Modular Pressure Controller Chassis, 5 Bay, BSP Manifold
8270A-5-NPT | Modular Pressure Controller Chassis, 5 Bay, NPT Manifold
PCM-STD-100M | High-Pressure Control Module, Standard Turndown, 100 MPa (15 000 psi)
PCM-STD-40M | Pressure Control Module, 827X, Standard Turndown, 40 MPa (6000 psi)
Accessories: 8270A and 8370A Modular High-Pressure Controllers / Calibrators
Other
CPS-100M | High-Pressure Contamination Prevention System, Std Port/Adptrs
CPS-40M-HC40 | Contamination Prevention System, 40 MPa (6000 psi), with HC40 test port and hand tight adaptors
SPLT-40M | Self-Purging Liquid Trap
PK-8270-BSP | Lines and Fittings Kit, 8270A BSP
PK-8270-NPT | Lines and Fittings Kit, 8270A NPT
PK-8270-SAE | Lines and Fittings Kit, 8270A 7/16-20
PK-8370-100M | Lines and Fittings Kit, 837X
TST-100M | High-Pressure Test Station, Std Port/Adaptors
GBK-110M | Gas Booster Kit, 152:1, 110 MPA (16 000 PSI)
GBK-50M | Gas Booster Kit, 75:1, 50 MPA (7300 PSI)
VA-PPC/MPC-REF-110 | Reference Vacuum Pump Kit, 110 V
VA-PPC/MPC-REF-220 | Vacuum Pump Package, 220 V
CASE-XX70 | Reusable Transit Case for XX70
KIT-PMM-CAL-100M | High-Pressure Measurement Module Calibration Kit
KIT-PMM-CAL-40M | Pressure Measurement Module Calibration Kit
CDG-REF-1TORR
PK-VLV-ABORT-100M | Kit, High-Pressure Abort Valve
PK-VLV-ISO-100M | Kit, High-Pressure Isolation Valve
PK-VLV-ISO-40M | Kit, Pressure Isolation Valve
KIT-EMM300 | Electrical Measurement Module with Docking Station
DS70-KIT-EMM | Electrical Measurement Module Docking Station
EMM300 | Electrical Measurement Module
SK-8270-SERVICE | Seals Kit, 8270 Service
SK-8370-SERVICE | Seals Kit, 8370 Service
Kits
TST-40M-HC40 | Test Station, Std Port/Adaptors
Cases and holsters
CASE-PMM|Shipping Case, 3 PMM Modules
Adapters and cables
PK-PMM-ZERO |Interconnection Kit for zeroing of Absolute Mode PM500 Modules
6270-SYS-CBL | 6270A System Cable Kit
Rack mounts and hardware
RMK-XX70 |Rack mount Kit, 19 In Width, 3U
For more information, please click here
Manuals
- 6270A, 8270A, 8370A | Reference guide
- 8270A/8370A | Safety sheet
- 8270A/8370A | Operators manual
- 8270A/8370A | Product specification
- CPS-40M-HC40/CPS-100 | Instruction sheet
- GBK-50M/GBK-110M | Instruction sheet
- KIT-EMM300 | Instruction sheet
- PK-8270-X/PK-8370-X | Instruction sheet
- PK-VLV | Instruction sheet
- SPLT-40M | Instruction sheet
- TST-40M-HC40/TST-100M | Instruction sheet
Application notes