z

Comtest

Fluke 190 Series III ScopeMeter Portable Oscilloscope

Sku: FLK-190-II-SeriesI

Price: Price range: R132238,50 through R247238,50 incl Vat

ScopeMeter Test Tools combine rugged portability with the high performance of bench oscilloscopes.

Product Information

High performance portable oscilloscopes engineered for harsh environments

Fluke 190 Series III ScopeMeter® Test Tools are engineered to go where you go, and tackle just about any troubleshooting job along the way. These CAT III 1000 V/CAT IV 600V rated test tools combine rugged portability with the high performance of bench oscilloscopes to help you take on the challenges of installing, commissioning and maintaining industrial machinery, automation and process controls, and power conversion electronics with ease—from DC to 500 MHz.

Choose from two or four channel models with a wide range of bandwidth options. Fast sampling rates up to 5.0 GS/s, 200 ps resolution and deep memory of 10,000 samples per channel allow high-accuracy capture and display of waveform details, noise, and other disturbances. Perform timing or amplitude related measurements on three phases or three-axis control systems, or simply compare and contrast multiple test points in a circuit under test. Features like TrendPlot™ Paperless Recorder, ScopeRecord™ Mode, Connect-and-View™ Triggering and a unique 100-screen Replay function help you quickly diagnose issues to minimize repair costs and downtime. These features make the oscilloscopes easy to use especially when diagnosing the most difficult problems like complex waveforms, induced noise, intermittent events and signal fluctuations or drift.

  • Up to four independent floating isolated inputs, up to 1000 V
  • Up to 5 GS/s real time sampling (depending on model and channels used)
  • Deep memory: 10,000 points per trace waveform capture (scope mode)
  • CAT III 1000 V/CAT IV 600 V safety rated instrument for industrial environments
  • Up to seven hours of battery operation using BP291
  • Large, bright color display is easy to view in nearly any environment
  • Easy to store and view historical data and transfer to a PC via USB or Wifi
  • Convenient battery access door for quick battery swaps in the field
  • IP51 rating, dust and drip-proof
  • Connect-and-View triggering for intelligent, automatic triggering on fast, slow and even complex signals
  • Frequency spectrum using FFT-analysis
  • Automatic capture and REPLAY of 100 screens
  • ScopeRecord mode gives 30,000 points per input channel for low frequency signal analysis
  • TrendPlot Paperless Recorder mode with deep memory for long- term automatic measurements
  • 5,000 count DMM included in the 2-channel models

    Fluke 190III Series Application TextMeasure from mV to kV safely

    Independently isolated inputs allow you to make measurements in mixed circuits having different ground references reducing the risk of accidental short circuits. Conventional bench oscilloscopes without special differential probes and isolation transformers can only reference measurements to line power earth ground. ScopeMeter 190 Series III test tools are engineered to cover a wide application range from mV to kV, so you’re ready for anything from microelectronics to heavy-duty higher voltage electrical applications. 190 Series III 60MHz and 100MHz configurations include VPS421 100:1 probes for higher voltage applications, while the 200MHz and 500MHz configurations include VPS410-II 10:1 probes suitable for both microelectronics and higher voltage applications.

    IP-51 rated for harsh environments

    Rugged and shock-proof, ScopeMeter Test Tools are built for dirty, hazardous environments. With its sealed case, it can endure dust, drips, humidity and airborne pollutants. Every time you reach for ScopeMeter Test Tool you can be confident it will work reliably wherever your work takes you.

    USB and Wi-Fi connectivity

    The Fluke 190 Series III offers two USB ports, electrically isolated from measurement input circuits allowing you to quickly and easily transfer data to a PC, archive and share waveforms with OEMs, colleagues and support staff, or store waveforms, screen captures and instrument setups onto USB memory devices for later use. Easily transfer saved files via USB stick, direct connection via the USB interface or optional Wi-Fi connectivity. These files can be used for further data handling or in FlukeView-2 Software to study waveforms in greater detail.

    Fluke 190III Series Screen 02Connect-and-View triggering

    Connect-and-View triggering provides an instant, stable display without the need for adjusting set- tings. If you’ve used other scopes, you know how tricky triggering can be. If settings are incorrect, results can be unstable or incorrect. Connect-and-View automatically sets up correct triggering by recognizing signal patterns. Without touching a button, you get a stable, reliable and repeatable display of virtually any signal including motor drive and control signals. It’s especially fast and convenient when you’re measuring a number of test points in rapid succession.

    Fluke 190III Series Screen 01Built-in digital multimeter

    Conveniently switch from waveform analysis to precise multimeter measurements using the built in 5000 count digital multimeter on two channel 190 Series III models. Measurement functions include Vdc, Vac, Vac+dc, resistance, continuity and diode test. Measure current and temperature using suitable shunt, probe or adapter with wide range of scaling factors.

    Fluke 190III Series Screen 04ScopeRecord™ mode for high resolution waveform recording

    ScopeRecord memory stores up to 30,000 or more data points per channel, capturing fast intermittent events and glitches as short as 8 ns. (Two sets of multiple channel recordings can be stored to internal memory for later analysis.)

    • Records events like UPS, power supply or motor start-up cycles
    • With the Stop on Trigger mode, the ScopeMeter Test Tool automatically recognizes a power failure and stores the waveform data preceding it

    Fluke 190III Series Screen 03

    The toughest faults to find are those that happen only once in a while. These intermittent events can be caused by bad connections, dust, dirt, corrosion, or simply broken wiring or connectors. Line outages, dips, swells and interruptions, or the starting and stopping of a motor can also cause a machine to stop. You may not be around when it happens, but the Fluke 190 Series III ScopeMeter Test Tool will be.

    • Plot minimum and maximum peak values and average over time
    • Plot any combination of up to four readings including voltages, amps, temperature, frequency and phase for all inputs, all with time and date stamp to pinpoint faults

     

    Fluke 190III Series Screen 05FlukeView™ 2 ScopeMeter software for documenting, archiving and analysisGet more out of your ScopeMeter Test Tool with FlukeView 2 ScopeMeter Software for Windows.

    • Documentation—transfer waveforms, screens and data to your PC for printing or importing data into a report
    • Add text to ScopeMeter Test Tool settings—give operators guidance when recalling settings
    • Archive—create a library of waveforms for easy reference, or waveform comparison
    • Analysis—use cursors or export data to another analysis program
  • Rated for industrial environments CAT III 1000 V/CAT IV 600 V
  • Automatically capture, view and analyze complex waveforms
  • Large, bright color display for easy in-the-field viewing
  • USB and Wi-Fi download for analyzing data with FlukeView® software
  • Up to four independent floating isolated inputs, up to 1000 V
Oscilloscope Models
  190-062

 

190-102

 

190-202

 

190-502

 

190-104

 

190-204

 

190-504

 

Vertical deflection

 

Number of channels

 

2 2 2 2 4 4 4
Bandwidth

 

60 MHz

 

100 MHz

 

200 MHz

 

500 MHz

 

100 MHz

 

200 MHz

 

500 MHz

 

Rise time

 

5.8 ns

 

3.5 ns

 

1.7 ns

 

0.7 ns

 

3.5 ns

 

1.7 ns

 

0.7 ns

 

Number of scope inputs

 

2 input channels plus external trigger

 

4 input channels

 

Channel architecture

 

All inputs fully insulated from each other and from ground. Inputs may be activated in any combination.

 

            Input coupling

 

AC or DC, with ground level indicator

 

Input sensitivity

 

With 10:1  probe, 20 mV  to 1000 V/div
With 100:1  probe, 200 mV to 10 kV/div
Direct (1:1),  2 mV  to 100 V/div

 

Bandwidth limiter

 

20 MHz and 10  kHz

 

Polarity

 

Normal,  Inverted, Variable

 

Input voltage

 

CAT III 1000 V/CAT IV 600 V rated, see General specifications for further details

 

Vertical resolution

 

8 bit

 

Accuracy at 4 s to 10 μs/div

 

5 mV/div to 100 V/div, ±(1.5 % + 6 counts)
2 mV/div, ±(1.5  % + 10 counts)

 

Input impedance

 

1 MΩ (± 1 %) //  15  pF (± 2.25  pF)

 

Horizontal
Maximum real- time sample rate (sampled simultaneously)

 

625 MS/s
(each channel)

 

1.25 GS/s (each channel)

 

2.5 GS/s (each channel)

 

5 GS/s (single channel) or
2.5 GS/s (dual channel)

 

1.25 GS/s (each channel)

 

2.5 GS/s (2ch)
1.25 GS/s (4ch)

 

5 GS/s (single ch) or 2.5 GS/s (2ch) or 1.25 GS/s (4ch)

 

Record length

 

Up to 10,000 samples per  channel

 

Time base range

 

10 ns/div
to 4 s/div

 

5 ns/div
to 4 s/div

 

2 ns/div
to 4 s/div

 

1 ns/div
to 4 s/div

 

5 ns/div
to 4 s/div

 

2 ns/div
to 4 s/div

 

1 ns/div
to 4 s/div

 

Time base in a 1-2-4-sequence
Slower time/division settings using ScopeRecord™ Roll mode (see ‘Recorder mode’)

 

            Maximum record length

 

10,000 samples per channel in scope mode
30,000 points per channel in ScopeRecord™ Roll mode (see ‘Recorder mode’)

 

Timing accuracy

 

± (0.01 % of reading + 1  pixel)

 

Glitch capture

 

8 ns (10  µs/div to 2 min/div)

 

Display and acquisition

 

Display

 

133 mm x 90 mm (5.3 in x 3.5 in) full-color high brightness LCD

 

Display modes

 

Any combination of channels; average on/off; replay.

 

Visible screen width

 

12  divisions horizontally in scope mode

 

Digital persistence modes

 

Off, short, medium, long, infinite and envelope mode

 

Waveform mathematics

 

One (190-xx2) or two (190-x04) mathematical operations on 2 input channels (A and B, C and D): add, subtract, multiply; X-Y-mode; Frequency Spectrum using FFT

 

Acquisition modes

 

Normal, Averaged, Auto, Single Shot, ScopeRecord™ roll, glitch capture, waveform compare with automatic “Pass/Fail testing”; Replay

 

Trigger and delay

 

Source

 

Input A, B or External (via meter input)

 

Input A, B, C or  D

 

Modes

 

Automatic, Edge, Pulse Width, N-Cycle, External (190-xx2)

 

Connect-and- View™

 

Advanced automatic triggering that recognizes signal patterns, automatically sets up and continuously adjusts  triggering, time base and amplitude. Automatically displays stable waveforms of complex and dynamic signals like motor drive and control signals. Can be switched off if preferred.

 

Pulse width triggering (on channel A)

 

Pulse width qualified by time
Allows for triggeringt, =t, ≠ t, where t is selectable in minimum steps of 0.01 div or 50 ns

 

Time delay

 

1 full screen of pre-trigger view or up to 100 screens (=1,200 divisions) of post-trigger delay

 

Dual slope triggering

 

Triggers on both rising and falling edges alike

 

N-cycle triggering

 

Triggers on N-th occurrence of a trigger event; N to be set in the range 2 to 99

 

Automatic capture of 100 screens

 

When in oscilloscope mode, the instrument ALWAYS memorizes the last 100 screens—no specific user setup required. When an anomaly is seen, the REPLAY button can be pressed to review the full sequence of screen events over and over. Instrument can be set up for triggering on glitches or intermittent anomalies and will operate in “baby-sit” mode capturing 100 specified events.

 

Replay

 

Manual or continuous replay. Displays the captured 100 screens as a “live” animation, or under manual control. Each screen has date and time-stamp.

 

Replay storage

 

Ten sets of 100 screens each can be saved internally for later recall and analysis. Direct storage of additional sets on external flash memory drive through USB host port.

 

FFT—frequency spectrum analysis

 

Shows frequency content of oscilloscope waveform using Fast Fourier Transform

 

Window

 

Automatic, Hamming, Hanning or None

 

Automatic window

 

Digitally re-samples acquired waveform to get optimum frequency resolution in FFT resultant.

 

Vertical scale

 

Linear/Logarithmic (in volts or amps)

 

Frequency axis

 

Frequency range automatically set as a function of timebase range of oscilloscope

 

Waveform compare and pass/fail testing

 

Waveform compare

 

Provides storage and display of a reference waveform for visual comparison with newly acquired waveforms. Reference is derived from an acquired waveform and can be modified in the oscilloscope.

 

Pass/Fail testing

 

In waveform compare mode, the oscilloscope can be set up to store only matching (“Pass”) or only non-matching (“Fail”) acquired waveforms in the replay memory bank for further analysis.

 

Automatic scope measurements

 

V dc, V ac rms, V ac+dc, Vpeak max, Vpeak min, Vpeak to peak, A ac, A dc, A ac+dc, frequency (in Hz), rise time (using cursors), fall time (using cursors), Power Factor (PF), Watts, VA, VA  reactive, phase (between 2 inputs A&B or C&D), pulse width (pos./neg.), duty cycle (pos./ neg.), temperature °C, temperature °F (not for Japan), dBV, dBm into 50 Ω and 600 Ω, VPWM ac and VPWM(ac+dc) for measurement on pulse width modulated motor drives and frequency inverters, V/Hz ratio;

 

Advanced power and motor drive functions

 

V/Hz ratio, Power Factor (PF), Watts, VA, VA reactive, V-PWM (ac) and V-PWM (ac+dc) for measure- ment on pulsewidth modulated motordrives and frequency inverter

 

Cursor measurements

 

Source

 

On any input waveform or on mathematical resultant waveform (excl. X-Y-mode)

 

Dual horizontal lines

 

Voltage at cursor 1 and at cursor 2, voltage between cursors

 

Dual vertical lines

 

Time between cursors, 1/T between cursors (in Hz), voltage between markers, risetime with markers, falltime with markers; Vrms between cursors, Watts between cursors.

 

Single vertical line

 

Min-Max and Average voltage at cursor position; frequency and rms-value of individual frequency component in the FFT Resultant

 

Advanced functions

 

mA*s (current-over-time, between cursors); V*s (voltage-over-time, between cursors); W*s (energy, between cursors)

 

ZOOM

 

Ranges from full record overview to zoom in up to sample level, at any record length.

 

Meter modes

 

  190-062

 

190-102

 

190-202

 

190-502

 

190-104

 

 

190-204

 

190-504

 

Meter inputs

 

Via 4 mm banana inputs, fully isolated from scope inputs and scope ground

 

Via BNC scope inputs

 

            Number of readings

 

One at a time via DMM input

 

Up to 4 automatic scope measurements simultaneously

 

Maximum resolution

 

5,000 counts

 

± 999 counts
(frequency: 9999 counts)

 

Input impedance

 

 

1 MΩ (± 1 %) // 14 pF (± 1.5 pF)

 

1 MΩ (± 1 %) // 15 pF (± 2.25 pF)

 

Advanced meter functions

 

Auto/manual ranging, relative measurements (Zero reference), TrendPlot™ recording

 

  The specified accuracy is valid over the temperature range 18 °C to 28 °C Add 10 % of specified accuracy for each degree C below 18 °C or above 28 °C

 

Voltage

 

V dc accuracy

 

± (0.5 % + 6 counts)

 

± (1.5 % + 6 counts)

 

V ac true rms accuracy

 

15 Hz to 60 Hz

 

± (1 % + 10 counts)

 

± (1.5 % + 10 counts)

 

60 Hz to 1 kHz

 

± (2.5 % + 15 counts)

 

 

60 Hz to 20 kHz

 

 

± (2.5 % + 15 counts)

 

Voltmeter ranges

 

 

500 mV, 5 V, 50 V, 500 V, 1,100 V

 

 
Resistance

 

Ranges

 

500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ

 

 

Accuracy

 

± (0.6 % + 6 counts)

 

 

Other meter functions
Continuity Beeper on < 50 Ω (± 30 Ω)
Diode test Up to 2.8 V
Current (A)

 

A dc, A ac, A ac+dc using an optional current clamp or shunt Scaling factors: 0.1 mV/A, 1 mV/A to 100 V/A and 400 mV/A
Temperature

 

With optional accessories. Scale factors 1mV/°C or 1mV/°F

 

Recorder mode
  190-062

 

 

190-102

 

190-202

 

190-502

 

190-104

 

190-204

 

190-504

 

ScopeRecord™ Roll Mode
Dual or multiple input waveform storage mode, using deep memory
Source and display

 

Input A, Input B,  Dual
All channels sampled simultaneously

 

Any combination of inputs, up to 4 channels.
All channels sampled simultaneously

 

Memory depth

 

30,000 data points per channel, each holding min/max pair of information

 

Min/max values

 

Min/max values are created at samples that are measured at high sample rate ensuring capture and display of glitches.

 

Recording modes

 

Single sweep, continuous roll; Start-on-Trigger (through external); Stop-on-Trigger (through external)

 

Single sweep, continuous roll; Start-on-Trigger (through any channel); Stop-on-Trigger (through any channel)

 

Stop-on-trigger

 

ScopeRecord mode can be stopped by an individual trigger event, or by an interruption
of a repetitive trigger signal, through any input channel (through External on 190-XX2 Series)

 

Horizontal scale

 

Time from start, time of  day

 

Zoom

 

Ranges from full record overview to zoom in up to sample level

 

Memory

 

Two multiple input ScopeRecord waveforms can be saved internally for later recall and analysis.

 

ScopeRecord™ Roll mode sample rate and recording timespan

 

Time base range

 

4 ms/div to 2 min/div

 

Recorded timespan

 

4.8 sec to 40 hr

 

Time/division in ‘view all’ mode

 

0.4 s/div to 4 h/div

 

Glitch capture

 

8 ns

 

Sample rate

 

125 MS/s

 

Resolution

 

160 μsec ~ 4.8 sec

 

Trendplot™ Recording

 

Multiple channel electronic paperless recorder. Graphically plots, displays and stores results of up to four automatic scope measurements or a DMM-reading over time.

 

Source and display

 

Any combination of scope measurements, made on any of the input channels, or DMM reading (2-channel instruments)

 

Memory depth

 

19,200 points (sets) per recording. Each recorded sample point contains a minimum, a maximum and an average value, plus a date- and time-stamp.

 

Ranges

 

Normal view: 5 s/div to 30 min/div; In view-all mode: 5 min/div to 48 hr/div (overview of total record)

 

Recorded time span

 

Up to 22 days, with a resolution of 102 seconds; up to 5.5 days for 4 readings.

 

Recording mode

 

Continuous recording, starting at 5 s/div with automatic time-scale compression

 

Measurement speed

 

Three automatic measurements per second or more

 

Horizontal scale

 

Time from start, time of day

 

Zoom

 

Up to 64x zoom-out for full record overview, up to 10x zoom-in for maximum detail

 

Memory

 

Two multiple input TrendPlot records can be saved internally for later recall and analysis.

 

Cursor measurements—all recorder modes

 

Source

 

            Any waveform trace in any waveform display mode (Scope, ScopeRecord or TrendPlot)

 

Dual vertical lines

 

Cursors may be used to identify Min, Max or Average value of any datapoint in a record, with time between cursors, time from start or absolute time.

 

General specifications

 

            Input voltage range

 

Rated maximum floating voltage

 

CAT III 1000 V / CAT IV 600 V (maximum voltage between any contact and earth-ground voltage level)

 

Probe input voltage VPS410-II

 

CAT III 1000 V / CAT IV 600 V (Maximum voltage between standard 10:1 probe tip and reference lead)

 

Probe input voltage VPS421

 

CAT III 1000V / CAT IV 600V (Maximum voltage between probe tip or reference lead to GND, 2000V max between probe tip and reference lead)

 

Maximum BNC input voltage

 

CAT IV 300 V (maximum voltage on BNC input directly)

 

Maximum volt- age on meter input

 

CAT III 1000 V / CAT IV 600 V
(safety designed banana input connectors)

 

 

Memory save and recall

 

Memory loca- tions (internal)

 

30 waveform memories plus 10 recording memories plus 9 screen copy memories

 

30 waveform memories

 

Each memory can contain up to 2 or 4 waveforms plus corresponding setups.

 

10 recording memories

 

Each may contain: a 100 Screen Replay sequence, or a ScopeRecord Roll-mode recording (2 or 4 traces), or a TrendPlot recording of up to 4 measurements

 

External data storage

 

On PC, using FlukeView™-2 Software, or direct storage on external flash memory drive (maximum 32 GB) through USB host port

 

Screencopies

 

On PC, using FlukeView™-2  Software, or internally (in instrument) which can be copied on to external flash memory drive as .BMP-file, through USB host port

 

Volatility

 

Saving is done in non-volatile Flash-ROM and all data is secured, independent of battery or power status.

 

Case
Design

 

Rugged, shock-proof with integrated protective holster. Handstrap and hangstrap included as standard.
Kensington lock supported to lock down instrument when left unattended.

 

Drip and dust proof

 

IP 51  according to IEC60529

 

Shock and vibration

 

Shock 30 g, vibration (sinusoidal) 3 g / 0.03 g2/Hz (Random), according to MIL-PRF-28800F Class 2

 

Display size

 

133 mm x 90 mm (5.3 in x 3.5 in) LCD

 

Resolution

 

1120  pixels x 765 pixels

 

Brightness

 

User-adjustable, up to 300  cd/m2

 

Mechanical data

 

Size

 

265 mm x 192 mm x 70 mm (10.5 in x 7.6 in x 2.8 in)

 

Weight (including battery)

 

2.1 kg (4.6 lb)

 

2.2 kg (4.8 lb)

 

Power

 

Line power

 

Universal mains adapter/battery charger BC190/830 included, with detachable 2-wire power cords 100 Vac to 240 Vac, ±10 %, 50-60 Hz

 

Battery power

 

Re-chargeable Li-Ion battery (included). Battery swappable through easily accessible battery door at the rear of the instrument

 

Battery type (incl.) and capacity [+opt. battery]

 

BP290: 10.8V, 2500 mAh
[BP291 (5000 mAh) optional]

 

BP291: 10.8V, 5000 mAh

 

Battery charge indicator Battery has built-in status indicator for use with external charger, next to battery status indicator on instrument screen.

 

Battery operat- ing time (with backlight low)

 

Up to 3.5 using BP290  (included), up to 7 hours using BP291 (optional)

 

 

Up to 7 hours using BP291 (included)

 

Battery charging time

 

2½ hours using BP290; 5 hours using BP291

 

Five hours for BP291

 

Battery power saving functions

 

Auto ‘power down’ with adjustable power down time. Automatic ‘display off’ with adjustable power down time.
On-screen battery power indicator

 

Safety

 

Compliance

 

EN61010-1,  Pollution Degree 2;
IEC 61010-2-030: CAT IV 600 V / CAT III 1000 V

 

Environmental

 

Operating temperature

 

Battery discharging: 0 °C to 40 °C (32 °F to 104 °F)
Battery charging: 0 °C to 40 °C (32 °F to 104 °F)

 

Storage temperature

 

-20  °C to 60 °C (-4 °F to 140  °F)

 

Humidity

 

0 °C to 10 °C (32 °F to 50 °F): noncondensing
10 °C to 30 °C (50 °F to 86 °F): 95 % (±5 %)
30 °C to 40 °C (86 °F to 104 °F): 75 % (±5 %)
40 °C to 50 °C (104 °F to 122 °F): 45 % (±5 %)

 

Maximum operating altitude

 

CAT IV 600 V, CAT III 1000 V: up to 2000 m (6 600 feet)
CAT IV 300 V, CAT III 600 V, CAT II 1000 V: up to 4000 m (13 000 feet)

 

Maximum storage altitude

 

12  km (40,000 ft)

 

Electro-Magnetic Compatibility (EMC)

 

IEC 61326-1: Industrial;
CISPR 11:  Group 1,  Class A;
Korea (KCC): Class A Equipment (Industrial Broadcasting and Communication Equipment);
USA (FCC): 47 CFR 15 subpart C.

 

Interfaces

 

Two USB-ports provided. Ports are fully insulated from instrument’s floating measurement circuitry.
USB-host port directly connects to external flash memory drive (up to 32 GB) for storage of waveform data, measurement results, instrument settings and screen copies. Alternatively, this USB-A port may be used to connect a WiFi Adapter for wireless PC connectivity. A mini-USB-B is provided which allows for interconnection to PC for remote control and data transfer under PC-control using FlukeView-2.

 

Probe calibration output

 

Dedicated probe-cal output with reference contact provided, fully insulated from any measurement input channel.
Generator Output: 1.225 Vpp / 500 Hz square wave

 

Warranty

 

3 years on main instrument, 1 year on battery and accessories

 

Included accessories

 

Battery charger/ mains adapter

 

BC190/830

 

Li-Ion battery pack

 

BP290 (10.8V, 2500 mAh)

 

BP291 (10.8V, 5000 mAh)

 

Voltage probe sets. Each set includes ground lead, hook clip; ground spring and probe tip insulation sleeve with VPS410-II-x.

 

 

2 pcs VPS421-x, ruggedized industrial-grade probes, 100:1, 150MHz with shrouded 4mm banana tip and large jaw alligator clips (one red, one blue)

 

 

2 pcs VPS410-II-x, 10:1 voltage probes, 500 MHz, (one red, one blue)

 

4 pcs VPS421-x, ruggedized probes, 100:1, 150 MHz, (red, blue, grey, green)

 

4 pcs VPS410-II-x, 10:1 voltage probes, 500 MHz, (one red, one grey, one blue, one green)

 

Test leads

 

TL175 (one red, one black) with test pins

 

 

Other

 

Li-Ion battery (BP290 or BP291,  see above), Battery charger (BC190) with universal power cord set, Hangstrap, Handstrap (user selectable for left- or right hand use), download information for user manual and FlukeView®-2 demo package (with restricted functionality), and USB interface cable for PC connectivity. Feedthrough cable terminator, 50 Ω (one per channel, 190-50x only).

 

Optional configuration

 

Each model is available as a ‘boxed’ version, described above, or with the optional SCC293 set included. SCC293 comprises: CXT293 rugged protective carrying case, full-version FlukeView PC software (activation code) and a WiFi dongle for wireless PC-connectivity using FlukeView-2 software.

 

Optional accessories

 

SCC293, VPS101 – 1:1 voltage probe; VPS510-x – wide bandwidth compact probes; i400s–current clamp; HH290–hanging hook; CXT293–protective carrying case; TRM50–BNC Feedthrough cable terminator, 50 Ω, safety designed; EBC290–battery charging bay

 

FLUKE-190-062-III 5282331 SCOPEMETER 2 CHANNEL 60MHZ Colour
FLUKE-190-062-III-S 5282354 SCOPEMETER 2 CHANNEL 60MHZ Colour with Software and Bag
FLUKE-190-102-III 5282368 SCOPEMETER 2 CHANNEL 100MHZ Colour
FLUKE-190-102-III-S 5282379 SCOPEMETER 2 CHANNEL 100MHZ Colour with Software and Bag
FLUKE-190-104-III 5282393 SCOPEMETER 4 CHANNEL 100MHZ Colour
FLUKE-190-104-III-S 5282417 SCOPEMETER 4 CHANNEL 100MHZ Colour with Software and Bag
FLUKE-190-202-III 5282439 SCOPEMETER 2 CHANNEL 200MHZ Colour
FLUKE-190-202-III-S 5282442 SCOPEMETER 2 CHANNEL 200MHZ Colour with Software and Bag
FLUKE-190-204-III 5282463 SCOPEMETER 4 CHANNEL 200MHZ Colour
FLUKE-190-204-III-S 5282474 SCOPEMETER 4 CHANNEL 200MHZ Colour with Software and Bag
FLUKE-190-502-III 5282495 SCOPEMETER 2 CHANNEL 500MHZ Colour
FLUKE-190-502-III-S 5282508 SCOPEMETER 2 CHANNEL 500MHZ COLOR SCC
FLUKE-190-504-III 5282513 SCOPEMETER 4 CHANNEL 500MHZ Colour
FLUKE-190-504-III-S 5282524 SCOPEMETER 4 CHANNEL 500MHZ COLOR SCC
Scroll to Top