The Fluke Calibration 2271A Industrial Pressure Calibrator provides a complete, automated pressure testing solution for calibrating a wide variety of pressure gauges and sensors. Thanks to its modular design, it can be configured to meet different needs and budgets, and expanded to cover a broad workload.

The 2271A is compatible with modules at two different accuracy classes, to provide maximum flexibility in workload and budget.  The PM200 modules provide 0.02% FS for most ranges (see specification tab for details).  The PM500 modules provide 0.01% Reading, allowing the 2271A to be used to test or calibrate higher accuracy transmitters and digital gauges.
 
The 2271A is ideal for calibration laboratories starting out in pressure calibration because it offers wide pressure measurement capabilities in a single instrument. Everything you need for calibrating pressure is included; just connect supply pressure. And your investment will stand the test of time: as your workload grows and changes, the 2271A can grow and change too. Just add measurement modules.
 
The 2271A automated pressure testing solution is also a great fit for labs that currently calibrate pressure transmitters and gauges and want to expand their capabilities or make their processes more efficient. This instrument is easy to set up and use, so you don’t have to reserve it for your most experienced technicians.  The built in electrical measurement and programmable tasks provide full automation, helping your lab run more efficiently.
 
If you are concerned about contamination from workload coming in from the field, the 2271A is a good choice for you as well. Its optional Contamination Prevention System (CPS) provides an important safeguard against that pervasive hazard.
 
A graphical user interface in your choice of ten languages and an intuitive menu structure, as well as an intuitive hardware design, make the 2271A remarkably easy to learn and to use.

  • Calibrate a wide range of gauges and sensors with a single instrument 
  • Two levels of accuracy, 0.01% reading or 0.02% FS
  • Wide measurement range from -100 kPa to 20 MPa (-15 to 3000 psi)
  • Removable pressure measurement modules make it easy to change or add measurement ranges
  • Integrated electrical measurement module with HART communication provides a complete solution for calibrating pressure transmitters
  • Built-in dual test ports enable you to connect multiple devices under test (DUTs)
  • Localized graphical user interface in choice of ten languages

Wide workload coverage for the present and the future

The 2271A features pressure ranges from -100 kPa to 20 MPa (-15 to 3000 psi), which covers the requirements of a wide range of gauges and sensors. Thanks to its modular design, the 2271A enables you to install two modules with different measurement ranges within the same chassis. You can purchase modules to match your current workload now; later, when your workload changes and grows, you can easily add ranges, swapping modules in and out according to your workload. This flexibility enables you to maintain your investment in the 2271A for years to come.

A built-in electrical measurement module (EMM) with HART capabilities enables you to perform closed loop, fully automated calibration on 4-20 mA devices such as smart transmitters, gauges and switches. Just set up the 2271A and then walk away to attend to other tasks.

The EMM supplies 24 V dc loop power for measuring mA and V dc. It has a built-in 250 Ohm resistor that can be toggled on or off, eliminating the need to have an external resistor to enable HART communications.

The 2271A accuracy specifications are provided in full and supported by a Technical Note that details its measurement uncertainty, so you know exactly what you are getting. The Technical Note is available for download on the flukecal.com website. As with all Fluke Calibration instruments, these specifications are conservative, complete and dependable.

Versatile pressure measurement modules

The 2271A uses PM200 and PM500 Pressure Measurement Modules that let you set up a system that matches your needs for accuracy and value:

Good: PM200 Pressure Measurement Modules Better: PM500 Pressure Measurement Modules
  • 0.02% FS specification makes it ideal for calibrating or testing pressure dial gauges, lower accuracy transmitters, or pressure switches
  • Rugged silicon pressure sensor design allows for faster control
  • Economic price point helps facilitate the purchase of backup modules, making sure you are never down for calibration
  • Highly characterized and linearized silicon pressure sensor provides an economical way of making accurate pressure measurements
  • 0.01 reading measurement uncertainty from 50% to 100% for most ranges allows for wide workload coverage
  • More than 45 ranges, from low differential pressures up to 20 MPa (3000 psi) to choose from. Your application is covered with this wide flexibility in configuration.

Install up to two pressure modules in the 2271A chassis at one time, mixing and matching module classes and ranges to get the combination that best suits your needs. There is no limit to the number of modules that can be used with the system, allowing you to change pressure ranges on the fly to meet your needs. Modules snap in and out quickly and easily through the front of the 2271A; just slide each into a specially-designed track and tighten the knob until you hear it click into place. The click tells you the module is safely in place; a special anti-torque guard on the knob prevents over-tightening so you never have to wonder if you tightened it too much or not enough.

Each module uses an enhanced face-seal design that has been leak tested to pressures three times higher than the maximum working pressure. You don’t have to worry about a leak in the system affecting your ability to measure and control pressure.

Conveniently located dual test ports and reference port

Dual test ports on the top of the 2271A let you easily mount two devices under test (DUTs). You can potentially double your throughput without spending time searching for fittings and tees. The vertical test ports let you easily connect analog dial gauges without the need for additional test stands or manifolds. Two test port types are supported, the standard HC20 or the P3000 test port. Both types of test ports enable you to make hand tight connections to traditional NPT, BSP, or metric pressure fittings. The HC20 test port includes easy grip features and integral support for M20 connections, whereas the P3000 style test port provides backwards compatibility for users of Fluke Calibration P3000 deadweight testers or P5500 pressure comparators.

A reference port is also located on top of the 2271A for applications that require an atmospheric reference.

Safety features protect operators and instruments

Each measurement module, as well as the main chassis, has pressure relief valves to protect the instrument and its operators from accidental overpressure. The 2271A has been designed using Sound Engineering Practices (SEP). With the internal relief valves, user-settable pressure limits, and emergency abort button, safety is the highest priority.

Preventing contamination

If your workload includes devices that contain different substances like water, oil and gas, you could be at risk for contamination—something getting into your system that isn’t supposed to be there. Contamination can clog a calibrator’s valves, wear out its parts, and make it difficult to maintain pressure. If the contamination gets into the sensor, it can actually change the calibrator’s behavior and throw off your readings. If contamination is a concern to you, order the optional Contamination Prevention System (CPS) to help keep the calibrator’s valves clean and free from debris.

The CPS provides an unprecedented level of protection by maintaining uni-directional flow away from the controller, a gravity sump system, and a two-stage filtering system.

Automated Pressure Testing: Automate with COMPASS® software for improved consistency and throughput

Fluke Calibration COMPASS for Pressure software is designed specifically for pressure calibration. It enables you to automate the 2271A and run complete automated pressure testing calibration sequences on single or multiple devices under test. COMPASS software removes the unknowns often associated with getting automated systems online. The 2271A also features a full remote interface that enables you to use it with custom software or other data acquisition equipment. Details about the interface are provided in the 2271A User Manual.

CarePlans help you manage cost of ownership

Reduce downtime and control your cost of ownership with a CarePlan. Fluke Calibration offers one-year, three-year and five-year Priority Gold CarePlans, which feature an annual standard or accredited calibration of your 6270A calibrator with guaranteed six-day in-house turnaround1 , plus free repairs with guaranteed ten-day in-house repair (includes calibration). One-year, three-year, and five-year Silver CarePlans are available for those customers who only want extended warranty coverage.

We’re here to help

Fluke Calibration’s testing, repair and calibration services are dedicated to filling your needs quickly and at a fair cost while maintaining the unmatched level of quality that is our trademark. Our pressure calibration laboratories are accreditedto ISO/IEC 17025. We maintain global calibration and repair facilities to help you keep your hardware in top working order.

General specifications
Mains
Power requirements 100 V ac to 240 V ac, 47 Hz to 63 Hz
Fuse T2A 250 V ac
Max power consumption 100 W
Environment
Operating Ambient Temperature Range 15 °C to 35 °C
Storage temperature -20 °C to 70 °C
Relative Humidity Operating: <80 % to 30 °C, <70 % to 35 °C Storage: <95 %, non-condensing. A power stabilization period of four days may be required after extended storage at high temperature and humidity.
Vibration MIL-T-28800D CLASS 3
Altitude (Operation) <2000 m
Warmup time 15 minutes after power up or module installation, when items previously stored within operating ambient temperature
Electromagnetic compatibility (EMC)
International IEC 61326-1: Controlled Electromagnetic Environment
  Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself. Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances. Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object.
Korea (KCC) Class A Equipment (Industrial Broadcasting & Communication Equipment) Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes.
USA (FCC) 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Compliance
Ingress protection IEC 60529: IP20
Safety IEC 61010-1, Installation Category II, Pollution degree 2
Dimensions and weights
Dimensions
Height 2271A-NPT-STD 305 mm (12 in)
2271A-BSP-STD 305 mm (12 in)
2271A-NPT-P3000 237 mm (9.33 in)
2271A-BSP-P3000 237 mm (9.33 in)
Width 442 mm (17.40 in)
Depth 446 mm (17.55 in)
Weight
Chassis only 15 kg (33.06 lbs)
Communication interfaces
Primary remote interfaces Ethernet, RS232, USB
Electrical Measurement Module (EMM)
Connection
Banana jack
Maximum 30 V dc w.r.t. chassis ground
Aux drivers 4 external Solenoid Drivers
24 V dc. 100 % duty cycle when turned, reducing to 40 % shortly after.

PM200 Modules Performance Specifications

PM500 Modules Performance Specifications

Operating characteristics
Control precision (Dynamic mode)
PM200-BG2.5K 0.005 % Range Span
PM500 <20 kPa full scale 0.002 % Range Span
All other ranges 0.001 % Range Span
Control turndown 10:1 (typical)
To meet the control specifications, supply pressure should not be greater than 10 times the range of the measurement module. Control turndown is defined as the relationship between the provided supply pressure and the appropriate supply pressure for the range.  For example, a unit with a 7 MPa (1000 psi) and 700 kPa range (100 psi) with a supply pressure of 7.7 MPa (1100 psi) provides control precision of 0.001 % range because 7 MPa is 10 times greater than 700 kPa.  A system with ranges of 20 MPa (3000 psi) and 700 kPa (100 psi) with supply pressure of 22 MPa (3300 psi) will have 0.001 % range control precision on the 20 MPa range but only 0.003 % control precision on the 700 kPa range. Control precision of 0.001 % on the low range can be achievedby reducing the supply pressure.
Low Control Point 1 kPa (0.15 psi) absolute
Settling Time (typical)
PM200-BG2.5K 40 seconds
All other PM200 ranges 20 seconds
PM500 £20 kPa Full Scale 45 seconds
PM500 >20 kPa Full Scale 30 seconds
Typical settling time is the time required to be within 0.005 % of the setpoint for 10 % steps into volumes of 0 to 50 cm3 and pressures above 50 kPa (7.25 psi) absolute. Lower absolute pressures require longer settling times depending upon quality of the vacuum pump, diameter and material of tubing used, and test volume.
Maximum Overshoot 0.02 % Range Span
Pressure Limits
Working Pressure Range -97.90 kPa (-14.2 psi) gauge to 20 MPa (3000 psi) gauge
Supply Port  23 MPa (3300 psi) gauge
Test Port 20 MPa (3000 psi) absolute
Reference Port  115 kPa (17 psi) absolute
Vent Port  150 kPa (22 psi) absolute
Relief Valves
Chassis Supply port relief valve is set to 24.1 MPa (-0/+700 kPa), 3500 psi (-0/+100 psi)
Exhaust port relief valve is set to ~830 kPa (120 psi).
Each Pressure Measurement Module (PMM) includes a module-specific pressure protection device.
Supply Gas Type
Clean Dry Air or Nitrogen (Industrial Grade, 99.5 %)
Maximum Particulate Contamination ≤1.25 micrometer (50 microinches)
Maximum Moisture Content -50 °C dew point
Maximum Hydrocarbon Content 30 ppm
Vacuum Supply
>50 liters per minute capacity with Auto Vent feature
Exhaust gas will pass through the vacuum supply.  Appropriate protections should be taken when operating at higher pressures.
Source
Loop Compliance Voltage ³24 V dc @ 20 mA (without 250 W built in resistor)  
  ³19 V dc @ 20 mA (with 250 W built in resistor)
Switch Testing
Current Limits £4 mA
Conditions Closed >1.5 mA, Open <0.5 mA
HART Communication and Functions
HART Modes HART mA measurement with 24 V (Loop)
Automatic HART Detection HART connected with automatic polling
HART Selectable Resistor Built-in 250 W loop resistor. Selectable ON/OFF
HART Commands HART universal and common practice commands (no device specific commands)
Write Protection HART Write enable/disable
Model Name Description
2271A

Industrial Pressure Calibrator

A complete pressure calibration system includes:

  • 1 chassis (2271A-NPT-XXX or 2271A-BSP-XXX)
  • 1 or more measurement modules (PM200 or PM500)
2271A-NPT-HC20

Industrial Pressure Calibrator Chassis, NPT Manifold, HC20 Test Port Connections

2271A-NPT-P3K

Industrial Pressure Calibrator Chassis, NPT Manifold, P3000 Test Port Connections

2271A-BSP-HC20

Industrial Pressure Calibrator Chassis, BSP Manifold, HC20 Test Port Connections

2271A-BSP-P3K

Industrial Pressure Calibrator Chassis, BSP Manifold, P3000 Test Port Connections

2271A-N-G20M

Pressure Calibrator, NPT, PM200-G20M, 20 MPA (3000 PSI)

2271A-N-G7M

Pressure Calibrator, NPT, PM200-G7M, 7 MPA (1000 PSI)

2271A-N-G2M

Pressure Calibrator, NPT, PM200-G2M, 2 MPA (300 PSI)

2271A-N-G20M/BG250K

Pressure Calibrator, NPT, PM200-G20M/BG250K, 20 MPA (3000 PSI)

2271A-N-G20M/G2M

Pressure Calibrator, NPT, PM200-G20M/G2M, 20 MPA (3000 PSI)

2271A-N-G2M/BG200K

Pressure Calibrator, NPT, PM200-G2M/BG200K, 2 MPA (300 PSI)

2271A-N-BG200K/BG2.5K

Pressure Calibrator, NPT, PM200-BG200K/BG2.5K, 200 KPA (30 PSI)

2271A-N-G7M/G700K

Pressure Calibrator, NPT, PM200-G7M/G700K, 7 MPA (1000 PSI)

2271A-N-G2M/A100K

Pressure Calibrator, NPT, PM200-G2M/A100K, 2 MPA (300 PSI)

2271A-N-G2M/BG100K

Pressure Calibrator, NPT, PM200-G2M/BG100K, 2 MPA (300 PSI)

Accessories common to all models:

Accessory Description
PM200 Pressure Measurement Modules

For 2271A Industrial Pressure Calibrator, and 6270A, 8270A  and 8370A Pressure Controller/Calibrators

PM500 Pressure Measurement Modules

For 2271A Industrial Pressure Calibrator and 6270A, 8270A and 8370A Pressure Controller / Calibrators 

CASE-2271

Shipping Case, 2271A

CASE-PMM

Shipping Case, 3 PMM Modules

CPS-2270-20M-P3K

Contamination Prevention System 20 MPa (3000 psi), P3000 Test Port Connection

CPS-2270-20M-HC20

Contamination Prevention System 20 MPa (3000 psi), HC20 Test Port Connection

PK-2271-NPT-HC20

Lines and Fittings Kit, 2271A NPT-HC20

PK-2271-NPT-P3K

Lines and Fittings Kit, 2271A-NPT-P3KPK

PK-2271-BSP-HC20

Lines and Fittings Kit, 2271A-BSP-HC20

PK-2271-BSP-P3K

Lines and Fittings Kit, 2271A-BSP-P3K

PMM-CAL-KIT-20M

Pressure Measurement Module Calibration Kit, 20 MPa (3000 psi)

VA-PPC/MPC-REF-220

Vacuum Pump Package, 220 V

VA-PPC/MPC-REF-110

Vacuum Pump Package, 110 V

CDG-REF-1TORR

Capacitance Diaphragm Gauge for zeroing of Absolute Mode PM500 Modules

PK-PMM-ZERO

Interconnection Kit for zeroing of Absolute Mode PM500 Modules

For more information, click here!

2271A Operators 

2271A Operators (chi) 

2271A Operators (fre) 

2271A Operators (ger) 

2271A Operators (ita) 

2271A Operators (jpn) 

2271A Operators (kor)

2271A Operators (por) 

2271A Operators (rus)

2271A Operators (spa) 

2271A Product Specifications 

2271A Product Specifications (chi)

2271A Product Specifications (fre) (2.43 MB)

2271A Product Specifications (ger) (2.83 MB)

2271A Product Specifications (ita) (2.42 MB)

2271A Product Specifications (jpn) (8.46 MB)

2271A Product Specifications (kor) (3.01 MB)

2271A Product Specifications (por) (2.82 MB)

2271A Product Specifications (rus) (3.67 MB)

2271A Product Specifications (spa) (2.82 MB)

2271A Programmers (613.76 KB)

2271A Safety Information (2.07 MB)

2271A Safety Information (chi) (1.91 MB)

2271A Safety Information (fre) (2.55 MB)

 2271A Safety Information (ger) 

2271A Safety Information (ita)

2271A Safety Information (jpn) 

2271A Safety Information (kor) 

2271A Safety Information (por) 

2271A Safety Information (rus)

2271A Safety Information (spa) 

Application notes

Automating the calibration of analog output pressure transmitters using the 2271A Industrial Pressure Calibrator 

Cardinal point calibration of pressure dial gauges using the Fluke 2271A Industrial Pressure Calibrator 

Guide to determining pressure measurement uncertainty for PM200, PM500 and PM600 Pressure Measurement Modules 

Simplifying the maintenance of HART pressure transmitters with the 2271A Industrial Pressure Calibrator 

data sheets

 2271A Industrial Pressure Calibrator Data Sheet 

Druckkalibrator 2271A für den Industrieeinsatz 

Étalonneur de pression industriel 2271A 

Calibrador de presión industrial 2271A 

2271A 一般産業用 圧力校正器 

Промышленный калибратор давления 2271A 

Calibrador de pressão industrial 2271A 

 2271A 산업용 압력 교정기 

2271A 工业 压力校准器 

2271A Industrial Pressure Calibrator Extended Specifications 

Firmware

2271A Embedded Software, Update by USB

video

2271A Industrial Pressure Calibrator Introduction Video